Products

Detector

Electrochemical Detector ED743

Electrochemical Detector ED743

Electrochemical Detector ED743

・Highly sensitive detection for electrochemically reactive
 compounds;
・Highly sensitive detection of sugars;
・Built-in online electrode cleaning function;
・Analog output function.

Specifications

Measurement method Amperometric, pulsed amperometric
Working electrode Diamond, gold, glassy carbon, platinum,
silver (select at time of purchase)
Reference electrode Silver‒silver chloride
Auxiliary electrode Titanium
Flow cell pressure resistance 1 MPa
Potential setting range ±5 V (10 mV steps)
Measurement range 10, 100, 1000 nA, 10, 100, 1000 μA/V
Response 0.1, 0.5, 1.0, 3.0, 6.0, 10.0 sec
Polarity switching Possible
Constant-temperature function 20‒45°C
Time program 30 steps, 10 files (stored in flash memory)
Input signal AUTO ZERO, START, STOP
Output signal Analog output (1 V, 10 mV), digital output (1 V, 10 mV),
EVENT, ERROR, READY
Size Main unit: 260 (W) × 420 (D) × 196 (H) mm
Constant temperature bath: 100 (W) × 300 (D) × 113 (H) mm
Weight Approximately 10 kg
Operating temperature 4‒35°C
Operating humidity 30‒80%
Power AC 100‒240 V, 50/60 Hz, 150 VA
Communication USB

DescriptionCat.No.
Electrochemical Detector ED743 Diamond6001-74310
Electrochemical Detector ED743 Gold6001-74311
Electrochemical Detector ED743 Glassy Carbon6001-74312
Electrochemical Detector ED743 Silver6001-74313
Electrochemical Detector ED743 Platinum6001-74314