Basics of Gas Chromatography
4-2 Baseline Problems
Baseline-related problems can range from column deterioration and contamination to equipment malfunctions. The baseline anomalies were divided into five categories: drift, spikes, noise, instability, and swell. Any problems related to equipment anomalies should be checked in the instruction manual of the equipment.
Drift
Main Causes
|
Approach
|
Insufficient column conditioning
|
Condition the column as described in Handling of Capillary Columns.
|
Air (water, oxygen) contamination
|
If excessive drift (baseline rise due to liquid phase stripping) occurs, check the carrier gas.
|
Degradation of column (liquid phase)
|
Replace the column.
|
Insufficient detector equilibration
|
Wait until the detector stabilizes.
|
Detector contamination
|
Clean the detector.
|
Spikes
Main Causes
|
Approach
|
Insufficient column connection
(detector side) |
This may occur when the column is inserted too far.
Check that the column insertion length matches the position specified in the instruction manual of the GC system. |
Detector contamination
|
Clean the detector.
|
Defective signal cable
|
Check for contact defects or insulation defects. Replace the cable if necessary.
|
Faulty electrical cable or power supply
|
Provide a stable power supply.
|
Noise
Main Causes
|
Approach
|
Detector contamination
|
Clean the detector.
|
Contamination of column (liquid phase)
|
Condition the columns. Cut approximately 50 cm off the inlet side of the column. If there is no improvement, replace the column.
|
Inlet contamination
|
This may be caused by liner contamination from inlet septum debris or sample residue. Clean or replace the liner. Replace the inlet septum.
|
Improper column connection (detector side)
|
The column might be inserted too far. Adjust the column insertion length to the position specified in the instruction manual of the GC system.
|
Inadequate carrier gas flow rate
|
Set the flow rate as specified in the instruction manual of the GC system.
|
Carrier gas leak (in MS, TCD, ECD)
|
May be caused by air leakage. Identify and remedy the leakage areas.
|
Detector failure
|
Replace or repair any defective filaments, electron multipliers, amplifiers, boards, etc.
|
Instability and Wander
Main Causes
|
Approach
|
Insufficient column conditioning
|
|
Contamination of column (liquid phase)
|
Condition the columns. Cut approximately 50 cm off the inlet side of the column. If there is no improvement, replace the column. |
Inlet contamination
|
This may be caused by liner contamination from inlet septum debris or sample residue.Clean or replace the liner. Replace the inlet septum. |
Detector contamination
|
Clean the detector.
|
Detector temperature instability (of TCDs)
|
Check the temperature stability of the device.
|
Unstable carrier gas flow rate, insufficient purity
|
Check the flow rate and gas purity (contamination).
|
Inappropriate temperature environment
around the equipment |
Check the ambient temperature.
|