Technical Information

Basics of Gas Chromatography

2-2 Capillary Column Selection Method

This section describes the points to consider when selecting a capillary column. These points are listed below.

  • Type of liquid phase
  • Column Length
  • Column I.D.
  • Film thickness of the liquid phase

The effects of each factor on the actual analysis are described below.

2-2-1 Types of Liquid Phases

Description Liquid Phase USP Code Polarity Application
InertCap 1MS 100% Dimethylpolysiloxane G2 None General Analysis, Hydrocarbon,
High Boiling Components, PCB, Phenol
InertCap 1 100% Dimethylpolysiloxane G2 None General Analysis, Hydrocarbon,
High Boiling Components, PCB, Phenol
InertCap 5MS/Sil 5% Diphenyl(equiv.)-
Dimethylpolysilphenylene siloxane
G27 Low General Analysis, Halogen Compound, Phenol, Pesticides, FAME
InertCap 5MS/NP 5% Diphenyl
95% Dimethylpolysiloxane
G27 Low General Analysis, Halogen Compound, Phenol, Pesticides, FAME
InertCap 5 5% Diphenyl
95% Dimethylpolysiloxane
G27 Low General Analysis, Halogen Compound, Phenol, Pesticides, FAME
InertCap Pesticides 5% Diphenyl(equiv.)
Dimethylpolysilphenylene siloxane
G27 Low Screening Analysis of Multiresidue Pesticides
InertCap 624 6% Cyanopropylphenyl
94% Dimethylpolysiloxane
G43 Middle VOC, Alcohol
InertCap 1301 6% Cyanopropylphenyl
94% Dimethylpolysiloxane
G43 Middle Pesticides, PCB, Alcohol, VOC
InertCap 25 25% Diphenyl
75% Dimethylpolysiloxane
G28 Middle Pesticides, PCB, Alcohol, VOC
InertCap 35 35% Diphenyl
65% Dimethylpolysiloxane
G42 Middle Pesticides, Amine, Medication, PCB
InertCap 1701 14% Cyanopropylphenyl
86% Dimethylpolysiloxane
G46 Middle Sugars, TMS derivatives, Drug, Alcohol, Steroid
InertCap 17MS 50% Diphenyl
50% Dimethylpolysiloxane
G3 Middle Steroid, Drug, Pesticides
InertCap 17 50% Diphenyl
50% Dimethylpolysiloxane
G3 Middle Steroid, Drug, Pesticides
InertCap 210 50% Trifluoropropyl
50% Methylpolysiloxane
G6 Middle Organophosphorus Pesticides
InertCap 225 50% Cyanopropylmethyl
50% Phenylmethylpolysiloxane
G19 Middle - High FAME
InertCap Pure-WAX Polyethylene Glycol G16 High General Analysis, Ester, Flavor, Alcohol, Aromatic, FAME
InertCap WAX Polyethylene Glycol G16 high General Analysis, Ester, Flavor, Alcohol, Aromatic, FAME
InertCap WAX-HT Polyethylene Glycol G16 High General Analysis, Ester, Flavor, Alcohol, Aromatic, FAME
InertCap FFAP Nitroterephthalic acid modified
Polyethylene Glycol
G35 High FAME, Free Fatty Acid, Organic Acid, Alcohol, Aldehyde

If the above columns (liquid phase) are unsuitable for the analysis or the analytical conditions should be further optimized, a dedicated column should be used.

Specific Columns

Description Liquid Phase USP Coed Liquid Phase Application
InertCap for Amines Special liquid phase Amine Compound, Alcohol
InertCap CHIRAMIX Special liquid phase Optical Isomer
InertCap AQUATIC 25% Diphenyl
75%Methylpolysiloxane
G28 Middle VOC, Organic Solvent
InertCap AQUATIC-2 25% Diphenyl
75%Methylpolysiloxanee
G28 Middle VOC, Organic Solvent

Liquid phases for general analyses are structurally divisible into silicon-based liquid phases and WAX-based liquid phases. Silicon-based liquid phases are characterized by the presence of methyl, phenyl, cyanopropyl, and other groups on the side chains of the Si–O–Si basic skeleton. Those with methyl groups on all side chains (called nonpolar liquid phases) are the least polar among all silicone-based liquid phases. Polar liquid phases are generated by replacing the methyl group by a phenyl or cyanopropyl group. The polarity of the liquid phase strengthens with increasing ratio of the polar group. The WAX-based liquid phase is highly polar and its basic backbone is a polymer of ethylene glycol. Free fatty acid phase (FFAP), a polymer of polyethylene glycol and nitro terephthalate, is even more polar and is used in analyses of free fatty acids and fatty acid esters.

Structural Formula of Silicon-Based Liquid Phase

Structural Formula of WAX-Based Liquid Phase

When selecting a liquid phase column, you must first match the polarities of the analyte and the liquid phase of the column. For example, if you are measuring polar substances such as alcohols or fatty acids, you should choose a WAX-based column or a highly polar FFAP column. If the polarity of the substance to be measured differs from that of the column, the solubility of the substance in the liquid phase might be reduced and the peak shape can be degraded. In addition, the sample load cannot be increased because an increased load reduces the allowable sample volume of the liquid phase. Below is an example of fatty acids measured on a nonpolar column and a highly polar column. The nonpolar column yielded poor peak shapes for the lower-molecular-weight fatty acids, whereas the highly polar column produced reasonable results. In the liquid phase of the highly polar column, the fatty acids spent more time dissolved in the liquid phase, causing stronger retention and slower elution times.

Comparison of Chromatograms from Different Liquid Phases

2-2-2 Column Length

The column length affects the theoretical plates and the analysis time. The lengths of commercial capillary columns range from 5–100 m, among which 30–60 m columns are commonly used. Doubling the length doubles the theoretical plates and the analysis time, but the separation improvement factor (~1.4) may be lower than expected. Longer columns are recommended for separation purposes whereas shorter columns are recommended for fast analyses. The selected column should be the shortest column that maintains the separation performance.

Deployment of Separation

R: resolution
N: number of theoreticalplates
α: separation factor
k: retention factor

2-2-3 Column I.D.

The inner diameter of the column affects the theoretical number of stages, degree of separation, and sample loading. In typical applications, the inner diameter is 0.25–0.53 mm. Columns with narrower inner diameters have a higher theoretical number of stages per unit length than wider columns, so are recommended when separation is important. However, narrow columns require enlargement of the split ratio or other adjustments to the introduction volume to compensate the small sample load. For trace analysis, narrow columns may be unsuitable and a column with a larger inner diameter (0.53 mm) and a small split ratio is recommended. Columns with narrow inner diameters (0.1–0.18 mm) are also suitable for fast analysis because their high resolution increases the linear velocity.

2-2-4 Film Thickness of the Liquid Phase

The thickness of the liquid phase film affects the analysis time and sample loading. A thick film allows higher sample loading and measurement of higher concentration samples than thin films. The chromatogram below shows how the separation improves on a thick film. In the column with 1.5 µm thick film, components other than propylene and propane were mostly separated. The retention ratio of each sample increased almost proportionally to the film thickness, lengthening the analysis time and widening the peak widths. When separating substances with high-boiling points, these broad peaks offset the separation benefits of increasing the film thickness. Moreover, thicker films are associated with larger baseline drift during programable temperature analyses.

2-2-5 Summary

We have discussed four factors that influence capillary column selection (liquid phase, length, internal diameter, and film thickness) and their effects on the analysis. Some more specialized options are mentioned below.

To find a specific column for a particular application, seek examples of previous analyses. If you find an example that matches your objective, you should be able to obtain a similar chromatogram under the analytical conditions described therein. If no analytical examples are found, the column should be selected after considering the four factors mentioned above. Next, the selected liquid phase should be chemically close to the target component. In many cases, either InertCap 1 or InertCap WAX will be a suitable choice. The length should be as short as possible without compromising the separation (a length around 30 m is usually recommended). The inner diameter should be 0.25 mm when separation is important and 0.53 mm when large amounts of sample are injected. The film should be as thin as possible within the analyzable range (a 0.25-µm-thick film is standard).

 

Many examples of analysis and know-how are available