Notification for Use of Columns
Maximum Operating Pressure, pH & Temperature Range
Description
|
Operating pH Range
(Normal Temperature)
|
Maximum Operating Temperature
|
|
InertSustain C18
InertSutain AQ-C18 InertSustainSwift C18 InertSustainBio C18 InertSustainSwift C8 |
1 - 10
|
60 ºC (pH1 - 7)
|
50 ºC (pH7 - 10)
|
InertSustain C8
InertSustain Phenylhexyl |
1 - 10
|
60 ºC (pH2 - 7)
|
50 ºC (pH7 - 9)
|
InertSustain Amide
|
2 - 8.5
|
60 ºC (pH2 - 8.5)
|
|
InertSustain Phenyl
InertSustain NH2 |
2 - 7.5
|
60 ºC (pH2 - 7.5)
|
|
Inertsil Series
|
2 - 7.5
|
60 ºC (pH2 - 7.5)
|
1) To prolong column life time, use them within pH range of eluent as above.
2) When using at pH 1-2 or pH 9-10, analysis at low temperature is recommended. The eluent containing organic solvent such as methanol is recommended.
3) The column lifetime varies greatly depending on conditions such as the pH, temperature, and solvent composition used.
- For analysis at pH 1-2, the use of TFA, formic acid, acetic acid, phosphate, etc. is recommended.
- It is recommended that organic buffers (e.g., triethylamine) of about 5 mM be used for the assay in pH 10.
- Use in the pH range of 2 to 8 when analyzing only buffer solution containing no organic solvent.